King Instrument Company
Home | About Us | Products | Flowmeter Sizing | Calibration Services | Meter Selection Guide | Contact Us
Call 714.891.0008
 
7430 Series - Glass
Both 65 mm and 150 mm scales, borosilicate glass tube models feature all stainless steel frame and horizontal
connections at a better price than competitive products. Valve optional.
 
Description
Metering Tube
Borosilicate Glass
Internal Components
316L Stainless Steel, Black Glass,
Sapphire, Carboloy, Tantalum
Inlet/Outlet Fittings
1/8" and 1/4" FNPT, Horizontal
ontrol Valve optional
Fitting Material
316L Stainless Steel
Elastomers
Standard: Viton
Optional: Buna N, EPR, and Kalrez®
Options
Alarm
Fiber-Optic or Inductive Ring Sensor
Certified Calibrations
Conform to ISA RP 16.6
Scales
Can be produced in any volumetric unit
Liquid Compatible
Gas Compatible
Performance
Capacities
.72 to 1,800 cc/min — Water
66 to 70,000 cc/min — Air
Scale
65 mm, 150 mm Direct reading, detachable
Accuracy
± 6% of Full Scale Flow, 65 mm
± 4% of Full Scale Flow, 150 mm
Turndown
10:1 to12.5:1 unless otherwise indicated
Repeatability .50%
Max Temperature
250° F (121° C) Gases Only
200° F (93° C) Liquids
Max Pressure
316L SS Fittings - 200 psig
PVC Fittings — 130 psig
PVDF Fittings — 150 psig
Ambient Temperature
33° F to 125° F (1° C to 52° C)

 

Materials Of Construction
Part Standard Optional
Metering Tube Borosilicate Glass
Frame 304 SS
Shield Polycarbonate
Fittings/Valve Assy. 316L SS
Float 316 SS or Black Glass Sapphire, Carboloy or Tantalum
O-Rings Viton® Buna-N, EPR, Kalrez®
Specifications
65MM Scale Flow Ranges
Tube
Number
Float
Material
Air (STP)
Water (70° F)
CC/MIN
SCFH
SLPH
CC/MIN
GPH
LPH
1-02-G-021 Glass 66 .14 4.0 0.72 .011 .042
Sapphire 105 .22 6.2 1.3 .021 .078
Stainless Steel 200 .42 12.0 3.3 .052 .190
Carboloy 340 .70 20.0 7.0 .110 .420
Tantalum 350 .74 21.0 7.8 .125 .460
1-03-G-041 Glass 76 .16 4.6 1.15 .018 .068
Sapphire 120 .25 7.2 2.10 .032 .125
Stainless Steel 230 .50 14.0 4.20 .068 .260
Carboloy 400 .85 24.0 9.00 .145 .560
Tantalum 440 .90 26.0 10.00 .165 .620
1-08-G-061 Glass 525 1.1 31 9.0 .14 .54
Sapphire 700 1.5 42 15.5 .24 .95
Stainless Steel 1130 2.4 68 29.0 .46 1.70
Carboloy 1600 3.4 95 46.0 .72 2.80
Tantalum 1700 3.6 100 50.0 .78 3.00
1-23-G-081 Glass 2000 4.4 120 44 .70 2.6
Sapphire 2600 5.4 150 68 1.05 4.0
Stainless Steel 3800 8.2 230 110 1.75 6.6
Carboloy 5600 12.0 340 170 2.70 10.5
Tantalum 6000 13.0 360 180 2.90 11.0
2-14-G-021 Glass 6800 14.5 400 160 2.6 9.5
Sapphire 9200 19.5 540 240 3.8 14.5
Stainless Steel 13000 28.0 800 400 6.2 24.0
Carboloy 19000 40.0 1100 600 9.5 36.0
Tantalum 20000 42.0 1200 640 10.0 38.0
2-34-G-041 Glass 19000 40.0 1150 520 8.20 31.0
Sapphire 25000 52.0 1500 740 11.50 44.0
Stainless Steel 42500 90.0 2550 1200 19.00 72.0
Carboloy 60000 125.0 3600 1700 27.00 105.0
Tantalum 70000 145.0 4200 1800 29.00 110.0
150 MM Scale Flow Ranges
Tube
Number
Float
Material
Air (STP)
Water (70° F)
CC/MIN
SCFH
SLPH
CC/MIN
GPH
LPH
1-03-G-022   Glass 54 .114 3.2 .56 .0088 .033
Sapphire 82 .175 4.9 1.04 .0160 .062
Stainless Steel 160 .340 9.8 2.25 .0350 .135
Carboloy 280 .580 16.5 5.00 .0780 .300
Tantalum 300 .620 17.5 5.20 .0840 .320
1-04-G-042 Glass 106 .225 6.4 1.24 .0195 .074
Sapphire 165 .35 10.0 2.35 .0380 .145
Stainless Steel 320 .68 19.0 5.60 .0900 .340
Carboloy 540 1.14 32.0 12.4 .1950 .740
Tantalum 580 1.24 35.0 13.5 .2100 .820
1-07-G-062 Glass 350 .74 21.0 4.7 .074 .28
Sapphire 500 1.06 30.0 10.0 .160 .60
Stainless Steel 820 1.75 50.0 20.5 .330 1.25
Carboloy 1250 2.60 76.0 34.0 .540 2.05
Tantalum 1350 2.90 80.0 36.0 .560 2.15
1-11-G-082 Glass 850 1.8 50.0 16.5 .26 1.0
Sapphire 1100 2.3 66.0 27.0 .42 1.6
Stainless Steel 1600 3.4 100.0 46.0 .72 2.7
Carboloy 2300 4.9 140.0 72.0 1.15 4.4
Tantalum 2450 5.2 155.0 80.0 1.25 4.8
1-27-G-102 Glass 2150 4.6 130.0 52 .84 3.1
Sapphire 2800 6.0 170.0 78 1.24 4.7
Stainless Steel 4400 9.2 260.0 130 2.05 7.8
Carboloy 6200 13.5 380.0 205 3.20 12.5
Tantalum 6750 14.0 400.0 210 3.30 12.5
2-09-G-002 Glass 3800 8.2 230.0 86 1.35 5.2
Sapphire 5000 10.6 300.0 130 2.05 7.8
Stainless Steel 7500 16.0 450.0 220 3.40 13.0
Carboloy 10600 22.5 640.0 330 5.20 20.0
Tantalum 11500 24.0 680.0 360 5.60 21.5
2-17-G-022 Glass 9000 19.0 540.0 215 3.40 13.0
Sapphire 11400 24.5 700.0 320 5.00 19.0
Stainless Steel 17000 36.0 1000.0 520 8.20 31.0
Carboloy 24000 50.0 1450.0 760 12.2 46.0
Tantalum 25000 54.0 1500.0 820 13.0 49.0
2-32-G-042 Glass 20500 43.0 1220.0 470 7.5 28.0
Sapphire 26000 56.0 1550.0 700 11.0 42.0
Stainless Steel 38000 82.0 2300.0 1120 18.0 68.0
Carboloy 54000 116.0 3300.0 1650 26.0 100.0
Tantalum 60000 125.0 3500.0 1750 28.0 106.0

Dimensions (Inches)

Detail
Letter
Scale length
65mm
150mm
A*
4.53
8.826
B
1.56
1.56
C
2.90
2.90
D
.73
.73
E
1.50
1.50
F
6.05
10.25
G
.50
.50
*The FNPT fittings have a 3/4 - 16 O.D. thread with mounting nuts installed.

 

Alarm Options:

Inductive Ring Sensor

Inductive ring sensors are designed to be used with a remote intrinsic safety barrier/switch isolator. These sensors are able to detect the metal float by producing an electromagnetic field within the ring. Ring sensors are available in either proximity or latching format for the 7400 Series.

Sensor Specifications
Power Supply: 5-25 VDC (from Switch Isolator)
Max Current, Target Present: 1 mA
Max Current, Target Absent: 15 mA
Operating Temperature Limits: 0°C to +40°C
Output: NAMUR
Repeatability: 0.01 mm
Switching Frequency: 2 kHz (.125”),
1.5 kHz (.25”)

Sensor Approvals:
UL Listed: General Purpose
FM Approved: Intrinsically Safe*
CSA Certified: Intrinsically Safe*
Cenelec: Intrinsically Safe*
*Additional cost, call for pricing
Float/Sensor Compatibility
Tube Float
Type Size Material
Proximity .125” SS, CB
Proximity .25” SS, CB
Latching .125” SS, CB
Latching .25” SS, CB

Fiber Optic Sensor
The fiber optic sensor is housed in a junction box attached to the side of a 7400 Series flowmeter. The sensor uses a pair of fiber optic cables, an emitter and receiver to transmit the light across the metering tube and back to the sensor. If the light beam is blocked by the float, the sensor output will change. The sensor provides a transistor output that switches the common or negative voltage (NPN) or positive voltage (PNP) to the load. The fiber optic sensor is compatible with all 7400 Series float types

Sensor Specifications
Supply Voltage: 10-30 Vdc
Current Consumption: 25 mA
Temperature Limits: -14°F to +212°F
Output:
NPN (Sinking), N.O. or N.C.
PNP (Sourcing), N.O. or N.C.
Offstate Leakage Current =1 microamp at 30 Vdc
Output Saturation Voltage = 1 V at 10 mA DC < 1.5 V at 150 mA DC

Note: Sapphire floats are not compatible with Fiber Optic Sensor



Ordering:

Use the following guide to determine the specific product number you require.
7
4
C
             
 
2
 
 
 
 
Meter
Series
Tube
Number
Float
Material
Fitting
Material
O-Ring
Material
Scale Valve
Option
Optional
Alarm Switch
See Specifications
Table
Glass - 1 316L SS
1/8" F NPT - 1
EPR - 1 Millimeter - 1 316L SS - Inlet - 1 Without Alarm - 0
  Sapphire - 2 316L SS
1/4" FNPT - 2
Buna-N - 2 GPH Water@70°F - 2 316L SS - Outlet - 2 Fiber-Optic NPN
(Proximity) - 1
316 SS - 3 PVC
1/4" FNPT - 3
Viton® - 3 LPH Water@70°F - 3 No Valve - 3 Fiber-Optic PNP
(Proximity) - 2
Carboloy - 4 PVC
1/8" FNPT - 4
Kalrez®
No Valve - 4
CC/MIN Water@STP - 4 PVC - Inlet - 4 Inductive Ring Sensor
(Proximity) - 3
Tantalum - 5 PVDF
1/4" FNPT - 5
Kalrez®
With Valve - 5
SCFH Air@STP - 5 PVC - Outlet - 5 Inductive Ring Sensor
(Latching) - 4
PVDF
1/8" FNPT - 6
SLPH Air@STP - 6 PVDF - Inlet - 6  
 Hastelloy® C
1/4" FNPT - 7
SCC/MIN Air@STP - 7
PVDF - Outlet - 7  
      Hastelloy® C
1/8" FNPT - 8
  Non standard - 8 Hastelloy® C
Inlet - 8
 
            Hastelloy® C
Outlet - 9
 
  Example: 74C - 102G021 - 1 - 2 - 3 -1 -1- 0
Ph: (714) 891-0008 • sales@kinginstrumentco.com • 12700 Pala Drive Garden Grove, CA 92841    |    Mission Statement   |    Quality Made In The U.S.A.